Testing the lightest and most insulating samples

To improve accuracy when analysing very low-conducting materials such as aerogels, this module is an essential addition. These materials displays such a low thermal conductivity that heat loss through the sensor medium – in this case cables – becomes significant.

Utilizing this module the heat loss can be accurately corrected for, making it possible to measure extremely low thermal conductivities.

The Low-Density / Highly-Insulating Measurement Module requires two sample halves for testing.

 

Thermal Conductivity Range:

Instruments Range
TPS 3500, TPS 2500 S, TPS 2200,
TPS 1500, TPS 1000
0.01 to 0.5 W/m/K